Kimo Temperature And Differential Pressure Transmitter – CP 215

Kimo Temperature And Differential Pressure Transmitter – CP 215

kimo-temperature-differential-pressure-transmitter-cp-215-9871034100

Product Specification for “Kimo Temperature And Differential Pressure Transmitter, CP 215”

Accuracy Range in Percent

±0.5 of reading ±2 mbar, (Pt100 temperature) : ±0.5 % of reading ±0.5 degreeC

Brand Name

Kimo View all products of Kimo

Display

19 digits 2 lines

Display Type

LCD

Frequency In Hz

50-60

Includes

LCC-S : configuration software with USB cable, SQR/3 function (square root for the measurement of air velocity and air flow), Calibration certificate, 115 Vac version transmitter, Connection tube, Connection fittings, Through-connections

Input Voltage (AC)

24 or 230, 115

Input Voltage in DC

24

Item Weight in grams

320 g

Material Type

ABS V0 as per UL94

Model

CP 215

Pressure Range

-2000/+2000 mbar

Product Information

Height of digits : Values : 10 mm ; Units : 5 mm
Pass-through : for cables Ø 6 mm maximum
Cable gland : for cables Ø 8 mm maximum
Output: 2 x 4-20 mA or 2 x 0-20 mA or 2 x 0-5 V or 2 x 0-10 V (4 wires), Common mode voltage <30 VAC, Maximum load : 500 Ohms (0/4-20 mA) / Minimum load : 1 K Ohms (0-5/10 V)
Galvanic isolation: Inputs and outputs (115 Vac/230 Vac models), Outputs (24 Vac/Vdc models)
Conditions of use (°C/%RH/m): From -10 to +50 degreeC. In non-condensing condition. From 0 to 2000 m
Please always use the device in accordance with its intended use and within parameters described in the technical features in order not to compromise the protection ensured by the device.

Protection

IP65

Resolution

1 mbar ; 1 mmH2O ; 0.1 kPa ; 0.1 inWG ; 1 mmHG ; 1 hPa ; 1 daPa ; 0.1 PSI

Standards

European directives: 2004/108/EC EMC ; 2006/95/EC Low Voltage ; 2011/65/EU RoHS II ; 2012/19/EU WEEE

Storage Temperature in Celsius

-10 to +70

TSIN

T009190540

Warranty

No Warranty

Key Features

Units of measurement: mbar, mmH2O, kPa, inWG, mmHG, hPa, daPa, PSI, °C / °F
Device fully protected by double isolation or reinforced isolation
Electrical connection: Screw terminal block for cables 2.5 mm2 . Carried out according to the code of good practice
PC communication: USB-Mini Din cable
Environment: Air and neutral gases
Autozero: Manual by push-button
Type of fluid: Air and neutral gases
Security: Protection class 2 ; Pollution degree 2 ; Overvoltage category 2
Configurable intermediate ranges, Trend indicator
¼ turn system mounting with wall-mount plate